2. Define the scope of the method to be validated and quantify the objectives
As an illustration, we will use the validation study of a method for determining nicotinamide (a vitamin B 3 vitamin) in milk by HPLC, with quantification by external calibration.
The natural content of nicotinamide in milk is between 0.6 and 1.0 mg L -
1
; regulations authorize supplementation, ranging from 0.4 to 2.0 mg L -
1
. The chosen validation range is therefore between 0.4 and 4.0 mg L -
1
.
There is no official reference for setting the acceptability limits of this method. However, the microbiological reference method has a high uncertainty of ± 40%. To demonstrate that the HPLC method can easily be an alternative to the reference method, acceptability limits...
You do not have access to this resource.
Exclusive to subscribers. 97% yet to be discovered!
Already subscribed?
Log in!
Ongoing reading
Define the scope of the method to be validated and quantify the objectives