6. Predicting recovered concentrations by inverse calibration
Before calculating the validation parameters, the instrumental responses measured on the samples (arbitrary unit) must be transformed into concentrations (mg L -1 ). This is done by taking the inverse function of the calibration model.
Note Y the instrumental response and X the concentration of the calibration solution. The calibration model is a straight line whose coefficients a 0 and a 1 can be estimated by the least-squares linear regression method : .
Once these coefficients a 0 and a 1 are known, the inverse calibration function used to predict the recovered concentration of...