2. Experimental procedure
2.1 Equipment
Most mercury porosimetry devices consist of three main components.
-
A pumping unit
It is used to evacuate the sample. Generally, a primary vacuum is achieved, i.e. a degassing pressure of less than 10 Pa.
-
A "low-pressure station" zone
The cell containing the sample is placed here. In this zone, the cell is evacuated, the mercury is introduced and the intrusion volume is analyzed as a function of pressure, down to a pressure close to atmospheric pressure. Mercury fills the cell spontaneously, as it is initially under vacuum. It fills the entire volume of the cell up to the entrance to the pores, where its pressure is...
Exclusive to subscribers. 97% yet to be discovered!
Already subscribed? Log in!
Experimental procedure
Article included in this offer
"Analysis and Characterization"
(
256 articles
)
Updated and enriched with articles validated by our scientific committees
A set of exclusive tools to complement the resources
Bibliography
- (1) - THOMMES (M.), KANEKO (K.), NEIMARK (A.V.), OLIVIER (J.P.), RODRIGUEZ-REINOSO (F.), ROUQUEROL (J.S.W.), SING (K.S.W.) - Physisorption of gases, with special reference to the evaluation of surface area and pore size distribution - (IUPAC Technical Report). Pure and Appl. Chem., 87(9-10), p. 1051-1069 (2015).
- ...