3. Direct line coupling, without splitter

The virtual disappearance of packed columns has rendered most molecular separators obsolete. Direct connection of a capillary column to a mass spectrometer is now the rule in all commercial GC-MS systems. However, even in the absence of a specific interfacing device, there are several ways of connecting a GC column directly to a mass spectrometer source, depending on the dimensions of the GC column, the vacuum pumping capabilities of the mass spectrometer, and the operating constraints specific to the laboratory (e.g. frequency of column exchange).

3.1 Sealed coupling

The coupling is tight when the end of the GC column is completely isolated from the laboratory atmosphere. There must be no air leakage at the point where the column is connected to the mass spectrometer,...

You do not have access to this resource.
Logo Techniques de l'Ingenieur

Exclusive to subscribers. 97% yet to be discovered!

You do not have access to this resource. Click here to request your free trial access!

Already subscribed?


Ongoing reading
Direct line coupling, without splitter

Article included in this offer

"Analysis and Characterization"

( 271 articles )

Complete knowledge base

Updated and enriched with articles validated by our scientific committees

Services

A set of exclusive tools to complement the resources

View offer details

Dans les ressources documentaires

Chromatographie ionique minérale - Méthodes de détection

Cet article se focalise sur la présentation des systèmes de détection couplés aux séparations des ions in...

Techniques analytiques pour les polluants émergents

Les polluants émergents sont des substances qui sont détectées dans l'environnement mais qui ne sont pas ...

Spectrométrie photoacoustique - Application à l’analyse de gaz

La méthode de la spectrométrie photoacoustiques (SPA) est l’objet principal de cet article. Cette méthode...

Tous les livres blancs
Toutes les actualités
Contact us